[关键词]
[摘要]
目的:对比分析Goldmann压平式眼压计(Goldmann applanation tonometer, GAT)和非接触眼压计(non-contact tonometer,NCT)测量青光眼患者眼压的结果,寻找其规律性。
方法:分别使用GAT和NCT测量108例206眼青光眼患者的眼压。
结果:青光眼患者108例206眼的平均眼压:GAT测量为29.77±10.27mmHg,NCT测量为24.59±8.58mmHg,两者比较存在显著性差异(P<0.01),而且眼压越高,差异性越显著。
结论:NCT测量值较GAT测量值低,眼压越高,差值越大。建议使用GAT检查青光眼患者的眼压,以免漏诊和贻误治疗。
[Key word]
[Abstract]
AIM: To compare the two methods for the measurement of glaucoma patients' intraocular pressure(IOP)between Goldmann applanation tonometer(GAT)and non-contact tonometer(NCT)and find the laws of the two methods.
METHODS: The IOP of 108 glaucoma patients(206 eyes)were measured by GAT and NCT respectively.
RESULTS: In 108 glaucoma patients, the average IOP of 206 eyes was 29.77±10.27mmHg by GAT and 24.59±8.58mmHg by NCT. There was significant difference between GAT and NCT(P<0.01). The higher of IOP, the difference between GAT and NCT was greater.
CONCLUSION: The measurement results with NCT were lower than that of GAT. The higher of IOP, the difference between GAT and NCT was greater. It's better to measure IOP by GAT for the glaucoma patients, in order to avoid the misdiagnosis and mistreatment of glaucoma.
[中图分类号]
[基金项目]