A new software for automated counting of glistenings in intraocular lenses in vivo
Author:
Corresponding Author:

Nick Stanojcic. Department of Ophthalmology, Guy’s and St Thomas’ NHS Foundation Trust, Westminster Bridge Road, London SE1 7EH, UK. nstanojcic@doctors.org.uk

  • Article
  • | |
  • Metrics
  • |
  • Reference [25]
  • |
  • Related [20]
  • | | |
  • Comments
    Abstract:

    AIM: To assess the performance of a bespoke software for automated counting of intraocular lens (IOL) glistenings in slit-lamp images. METHODS: IOL glistenings from slit-lamp-derived digital images were counted manually and automatically by the bespoke software. The images of one randomly selected eye from each of 34 participants were used as a training set to determine the threshold setting that gave the best agreement between manual and automatic grading. A second set of 63 images, selected using randomised stratified sampling from 290 images, were used for software validation. The images were obtained using a previously described protocol. Software-derived automated glistenings counts were compared to manual counts produced by three ophthalmologists. RESULTS: A threshold value of 140 was determined that minimised the total deviation in the number of glistenings for the 34 images in the training set. Using this threshold value, only slight agreement was found between automated software counts and manual expert counts for the validating set of 63 images (κ=0.104, 95%CI, 0.040-0.168). Ten images (15.9%) had glistenings counts that agreed between the software and manual counting. There were 49 images (77.8%) where the software overestimated the number of glistenings. CONCLUSION: The low levels of agreement show between an initial release of software used to automatically count glistenings in in vivo slit-lamp images and manual counting indicates that this is a non-trivial application. Iterative improvement involving a dialogue between software developers and experienced ophthalmologists is required to optimise agreement. The results suggest that validation of software is necessary for studies involving semi-automatic evaluation of glistenings.

    Reference
    1 Stanojcic N, Hull C, O’Brart DP. Clinical and material degradations of intraocular lenses: a review. Eur J Ophthalmol 2020;30(5):823-839.
    2 Tandogan T, Auffarth GU, Son HS, Merz P, Choi CY, Khoramnia R. In-vitro glistening formation in six different foldable hydrophobic intraocular lenses. BMC Ophthalmol 2021;21(1):126.
    3 Stanojcic N, OʼBrart D, Hull C, Wagh V, Azan E, Bhogal M, Robbie S, Li JO. Visual and refractive outcomes and glistenings occurrence after implantation of 2 hydrophobic acrylic aspheric monofocal IOLs. J Cataract Refract Surg 2020;46(7):986-994.
    4 Weindler JN, Łabuz G, Yildirim TM, Tandogan T, Khoramnia R, Auffarth GU. The impact of glistenings on the optical quality of a hydrophobic acrylic intraocular lens. J Cataract Refract Surg 2019;45(7):1020-1025.
    5 Wang Q, Yildirim TM, Schickhardt SK, Łabuz G, Khoramnia R, Merz PR, Son HS, Munro DJ, Friedmann E, Auffarth GU. Quantification of the in vitro predisposition to glistening formation in one manufacturer’s acrylic intraocular lenses made in different decades. Ophthalmol Ther 2021;10(1):165-174.
    6 Colin J, Orignac I. Glistenings on intraocular lenses in healthy eyes: effects and associations. J Refract Surg 2011;27(12):869-875.
    7 Johansson B. Glistenings, anterior/posterior capsular opacification and incidence of Nd: YAG laser treatments with two aspheric hydrophobic acrylic intraocular lenses - a long-term intra-individual study. Acta Ophthalmol 2017;95(7):671-677.
    8 Moreno-Montañés J, Alvarez A, Rodríguez-Conde R, Fernández-Hortelano A. Clinical factors related to the frequency and intensity of glistenings in AcrySof intraocular lenses. J Cataract Refract Surg 2003;29(10):1980-1984.
    9 Chang A, Behndig A, Rønbeck M, Kugelberg M. Comparison of posterior capsule opacification and glistenings with 2 hydrophobic acrylic intraocular lenses: 5- to 7-year follow-up. J Cataract Refract Surg 2013;39(5):694-698.
    10 Auffarth GU, Brézin A, Lignereux F, Khoramnia R, Yildirim TM, Kohnen T, Bianco J. Randomized multicenter trial to assess posterior capsule opacification and glistenings in two hydrophobic acrylic intraocular lenses. Sci Rep 2023;13(1):2822.
    11 House P, Abdul Rahman A, Richards J, et al. Long-term clinical audit of glistenings in Alcon Acrysof intra-ocular lenses with and without yellow chromophore. Clin Exp Ophthalmol 2020;48(2):251-252.
    12 Stanojcic N, O’Brart DPS, Maycock N, Hull CC. Effects of intraocular lens glistenings on visual function: a prospective study and presentation of a new glistenings grading methodology. BMJ Open Ophthalmol 2019;4(1):e000266.
    13 Kawai KJ. An evaluation of glistening and stability of intraocular lens material manufactured by different methods. Eur J Ophthalmol 2021;31(2):427-435.
    14 van der Mooren M, Safran S, Weeber HA, Piers P. Trend in Glistening density in Acrylic Intraocular Lenses and its relation to straylight performance. Invest Ophthalmol Vis Sci 2020;61:599.
    15 Kanclerz P, Grzybowski A. Glistenings might be associated with disability glare. Eur J Ophthalmol 2022;32(1):NP296.
    16 Stanojcic N, Hull C, O’Brart DPS. Reply to letter from Drs. Piotr Kanclerz and Andrzej Grzybowski entitled Glistenings might be associated with disability glare. Eur J Ophthalmol 2022;32(1): NP347-NP34愸爮攼⁢摲甾爱椷渠杂瑲桫敥䍳佴噥䥩䑮ⴠㅁ㥆‬瀠慂湯摲敫浥楮捳⹴⁥䅩獮椠慅⁍倮愠捐䩬⁡佲灩桺瑥桤愠汧浬潡汳⁳⡥偳栠業污慹⤠⁨⁥㉬ば㈠ㅩ㭮ㄠび⡹㑭⥰㩴㍭㡡ㅴ⵩㍣㠠㝣⹡㱳扥牳㸠㉯㥦†坩慮湴杲⁡呯Ᵽ⁵塬楡慲†䩬ⱥ䱳椠⁧剬奩ⱳ⁴敥瑮⁩慮汧⹳‮䤠湃瑬敩汮氠楏杰整湯瑭†挲愰琱愹爻愱挱琺‵猷甭父朲攮爼祢⁲猾由瀸攠牂癨楡獲楡潴湨⁩愠湍搬†敓癥慮汴畨慩瑬椠潓渮†癃楡慲⁴搭敷敨灥⁥汬攠慰牡湴楴湥杲⹮†䥩湮瑴⁲䩡卣畵牬条⁲㈠ぬ㉥㉮㭳ㄠで㑬㩩ㅳぴ㙥㝮㑩の⹧㰮戠爲㸰㌲〱‮䰼慢湲搾椱猹†䩓剡ⱹ䭯潲挠桄⁍䜬䜠⹃呬桥敭⁡浮攠慒獩畣牨敡浲敤湳瑯漠晄‬漠扄獡敩牲瘠敂牊‬愠材牯敬敬浡散湫琠⁓晋漮爠⁏捳慭瑯整杩潣爠楣捡慶汩⁴摡慴瑩慯⹮†䉯楦漠浥敬瑡牳楴捯獭⁥ㅲ㥩㝣㜠㭩㍮㍴⡲ㅡ⥯㩣ㅵ㕬㥡⵲ㄠ㝬㑥⹮ses. Acta Biomater 2010;6(3): 1090-1098.
    20 Philippaki E, O’Brart DP, Hull CC. Comparison of glistenings formation and their effect on forward light scatter between the Acrysof SN60WF and Eternity Natural Uni NW-60 intraocular lenses. BMJ Open Ophthalmol 2020;5(1):e000399.
    21 Xi L, Liu Y, Zhao F, Chen C, Cheng B. Analysis of glistenings in hydrophobic acrylic intraocular lenses on visual performance. Int J Ophthalmol 2014;7(3):446-451.
    22 Borghesi S, Colciago S, Zeri F, Scialdone A, Tavazzi S. In vitro glistening formation in IOLs: automated method for assessing the volumetric density and depth distribution of microvacuoles. J Cataract Refract Surg 2020;46(8):1178-1183.
    23 Yildirim TM, Fang H, Schickhardt SK, Wang Q, Merz PR, Auffarth GU. Glistening formation in a new hydrophobic acrylic intraocular lens. BMC Ophthalmol 2020;20(1):186.
    24 Yildirim TM, Schickhardt SK, Wang Q, Friedmann E, Khoramnia R, Auffarth GU. Quantitative evaluation of microvacuole formation in five intraocular lens models made of different hydrophobic materials. PLoS One 2021;16(4):e0250860.
    25 Schweitzer C, Orignac I, Praud D, Chatoux O, Colin J. Glistening in glaucomatous eyes: visual performances and risk factors. Acta Ophthalmol 2014;92(6):529-534.
    26 Jitpakdee P, Uyyanonvara B. Computer-aided detection and quantification in glistenings on intra-ocular lenses. Multimed Tools Appl 2017;76(18):18915-18928.
    27 Chen WB, Li RY, Yu QJ, et al. Early detection of visual impairment in young children using a smartphone-based deep learning system. Nat Med 2023;29(2):493-503.
    28 Kim SE, Logeswaran A, Kang S, Stanojcic N, Wickham L, Thomas P, Li JP O. Digital transformation in ophthalmic clinical c
    Cited by
    Comments
    Comments
    分享到微博
    Submit
Get Citation

Nick Stanojcic, Christopher C. Hull, Eduardo Mangieri,/et al.A new software for automated counting of glistenings in intraocular lenses in vivo. Int J Ophthalmol, 2023,16(8):1237-1242

Copy
Share
Article Metrics
  • Abstract:314
  • PDF: 621
  • HTML: 0
  • Cited by: 0
Publication History
  • Received:June 03,2022
  • Revised:May 24,2023
  • Online: July 25,2023